326 Chapter 10
CDMA Digital Modulation
cdma2000 Forward Link Modulation for Receiver Test
“Setting EbNo” on page 326
Setting the Carrier to Noise Ratio
1. Press
Mode Setup to return to the top-level real-time cdma2000 menu.
2. Press
Noise Setup > C/N > 10 > dB.
3. Press
Noise Off On to On.
You have now set the overall carrier to noise ratio to 10 dB and turned on noise. This is done to apply a
discernible level of noise across the entire channel space.
Setting EbNo
EbNo can be set in the Noise Setup menu or the Channel Setup table editor (see “Modifying Channel
Parameters” on page 323). This task teaches you how to make quick EbNo adjustments using the
EbNo
softkey in the Noise Setup menu.
1. Press
Channel Number and move the cursor to highlight the forward fundamental channel (F-FCH).
2. Press
EbNo > 20 > dB.
You have now set the EbNo value for the F-FCH channel to 20 dB. Note that modifying the EbNo value for
a channel will change the overall carrier to noise ratio and the EbNo value for all other active channels. For
reliable results when making hand calculations to determine or verify EbNo values, it is recommended that
you first scale the code domain power to 0 dB to display the normalized relative channel power levels (see
“Scaling to 0 dB” on page 324).
Generating the Baseband Signal
This procedure builds upon the previous procedure.
Press
CDMA2000 Off On to On.
This generates the real-time forward link cdma2000 signal. The signal generator may require several
seconds to build the signal. During this time, Baseband Reconfiguring appears on the display. After
the reconfiguration is complete, the display reads CDMA2000 On and the CDMA2K and I/Q annunciators
appear. The signal is now modulating the RF carrier.
The signal’s configuration data resides in volatile memory and is not recoverable after an instrument preset,
power cycle, or reconfigured signal generation.

Table of Contents

Other manuals for Agilent Technologies E4438C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4438C and is the answer not in the manual?

Agilent Technologies E4438C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4438C
CategoryPortable Generator
LanguageEnglish

Related product manuals