360 Chapter 11
GPS Modulation
Real Time GPS
Testing Receiver Sensitivity
Refer to Figure 11-8.
1. Connect the cables between the receiver and the ESG as shown in Figure 11-8.
Figure 11-8 Setup for a Receiver Sensitivity Test
2. Set the GPS data mode to TLM.
3. Set the power level on the ESG.
4. Set the L-band frequency on the ESG.
5. Set up the UE (user equipment) to receive the signal from the ESG.
6. Turn on the GPS personality and the RF output on the ESG.