EasyManua.ls Logo

Agilent Technologies E4438C

Agilent Technologies E4438C
700 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 279
BERT
Measuring RF Loopback BERT with Option 300
4. Press Pass/Fail Limits > Class II RBER > 2 > %.
5. Press
Return > Threshold # of Events to Stop > Class II Bit Error > 300 > Enter.
Notice that the
Class II Bit Error softkey is highlighted, 300 events is displayed beneath the softkey, and
Stop Thrs: CII(300) appears in the status area of the display.
Figure 9-10
6. Press the
Trigge r hardkey to start the measurement:
You will see Pass or Fail displayed on the lower left corner of the screen, when either of the following
occurs:
The measurement completes normally; in this case, after 100 frames.
The measurement terminates early, due to the designated number of events to stop being reached.
Refer to Figure 9-11 on page 280.

Table of Contents

Other manuals for Agilent Technologies E4438C

Related product manuals