How to Add a Reflective Event 4 Sample Sessions: Analyzing an Existing Trace
E6000C Mini-OTDR Users Guide, E0302
115
Reflection Height If you have chosen the Reflection Parameter to be
Reflection Height, reflectance is listed as Refl.H. at A
or Refl.H. at B.
See “How to Set the OTDR Settings” on page 134.
16 When you have positioned the level-markers correctly,
press S
ELECT to continue.
17 Press S
ELECT again to select Ok.
How to Set the Level-markers for
Measuring Insertion Loss
NOTE If you just want to measure the insertion loss, select
[Events]<Add/Mod. Non-Refl Event> and observe the
following steps.
See “How to Add a Non-Reflective Event” on page 119.
Level markers You now see 4 level-markers on the trace, and a message
asking you to adjust them (Figure 40). These four level-
markers allow you to analyze the Insertion Loss.
Two level-markers are to the left of the Event, and are
joined by a regression line. Two more level-markers are to
the right of the Event.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6000C and is the answer not in the manual?

Agilent Technologies E6000C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6000C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals