Loading...
Agilent E6006A Power Meter Submodule CSpecifications
E6000C Mini-OTDR Users Guide, E0302
215
Supplementary Performance
Characteristics
Automatic Zeroing Circuitry.
Automatic Ranging.
Modulation frequency recognition (270 Hz, 1 kHz,
2 kHz) is available at power levels between +10 and -
45 dBm (peak amplitude).
Specifications
Uncertainty at reference
conditions:
± 3%
Power level: -20 dBm
Continuous wave (CW)
Wavelength: 1300
±3 nm, 1310±3 nm, 1550±3nm
Fiber type: 50/125
µm graded index, Agilent/HMS-10
connector
Spectral bandwidth: up to 10 nm
Ambient temperature: +18 to +28
°C
At day of calibration (add 0.3% for aging of over one
year; add 0.6% for aging of over two years).
Tot al uncertainty:
±5% ± 0.5 nW (1310, 1550 nm)
Power level: +0 to -50 dBm
Continuous Wave (CW)
Wavelength: 850
±3 nm, 1300±3 nm, 1310±3nm,
1550
±3nm
Fiber type: SM to 50
µm graded index
(add 2% to total uncertainty for fiber 62.5
µm).
Straight and angled connectors
Ambient temperature: +10 to +40
°C
Within 2 years after calibration