D Single-Mode/Multimode Module Performance Tests Test V. E6006A Power Meter Submodule
258 E6000C Mini-OTDR Users Guide, E0302
Test V. E6006A Power Meter
Submodule
NOTE You may also use the following older, but discontinued,
equipment: 8153A (for 8163A), 81533B (for 81618A),
81524A #C01 (for 81624A #C01), and 81554SM with
81210LI #011 and 81310LI #011 (for 81654A).
Table 5 Power Meter: Test Equipment Required
Instrument or Accessories qty
Recommended
Agilent Model
Required Characteris-
tic
Alternative Models
Lightwave Multimeter In-
terface Module (1 channel)
Optical Detector Head
Laser Source 1310/1550
nm
Optical attenuator
Patchcord (HMS10/
HMS10, 9/50
µm, SM)
Patchcord (HMS10/
HMS10, 50/125 mm)
connector interface
connector adapter
1
1
1
1
1
1
4
1
8163A
81618A
81624A #C01
81654SM
8156A #101
81101AC
81501AC
81000AI
81000AA
1310 ± 10 nm
1550
± 10 nm
short term stability <
±0.005 dB
Attenuation > 50 dB
Return Loss > 45 dB
Repeatability <
±0.01 dB
8164A, 8166A
81619A
81657A
81650A and 81651A
81655A and 81656A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6000C and is the answer not in the manual?

Agilent Technologies E6000C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6000C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals