4 Sample Sessions: Analyzing an Existing Trace How to Calculate Reflectance or Insertion Loss for
122 E6000C Mini-OTDR Users Guide, E0302
Set level markers 3 Follow the steps in “How to Set the Level-Markers for
Measuring Reflectance” on page 112 to position the
level-markers properly.
4 Read the Reflectance for the Event in the Marker Info.
window.
The Reflectance is written at Refl. at A (or Refl.
at B, depending in the current marker).
NOTE If you have chosen the Reflection Parameter to be
Reflection Height, reflectance is listed as Refl.H. at A
or Refl.H. at B.
See “How to Set the OTDR Settings” on page 134.
Remove level markers 5 Remove the level-markers by deselecting
[A
NALYSIS]<ANALYZE REFLECTANCE>.
How to Calculate Insertion Loss
1 Move the active marker to an Event.
Analyze Insertion Loss 2 Select the [A
NALYSIS]<ANALYZE INSERTION LOSS> menu
option.
Set level markers 3 Follow the steps in “How to Set the Level-markers for
Measuring Insertion Loss” on page 115 to position the
level-markers properly.
4 Read the Insertion Loss for the Event in the Marker
Info. window.
The Insertion Loss is written at Ins. L. at A (or
Ins. L. at B, depending in the current marker).
Remove level markers 5 Remove the level-markers by deselecting
[A
NALYSIS]<ANALYZE INSERTION LOSS>.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6000C and is the answer not in the manual?

Agilent Technologies E6000C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6000C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals