D Single-Mode/Multimode Module Performance Tests Performance Test Form Sheets
248 E6000C Mini-OTDR Users Guide, E0302
Performance Test for the Agilent E6000C with Single-mode Modules Page 5 of 8
Model Agt. __________ Module Report No. __________ Date ___________
No. Test Description
I. Dynamic Range 1550 nm Wavelength
Minimum Specification
Pulsewidth Dynamic
Range
98%
Dyn Range =
Dyn Range
98%
+ 1.9dB
E6003A E6003B E6004A E6008B
E6012A
E6013A Meas.
Uncertainty
10
µs
1
µs
100 ns
10 ns
dB
_______
_______
_______
_______
dB
_______
_______
_______
_______
dB
34
29
22
17
dB
37
29
22
17
dB
28
23
18
13
dB
41
34
27
22
dB
35
28
22
17
dB
_______
_______
_______
_______
II. Event Deadzone 1550 nm Wavelength
Event Deadzone Max Spec Meas.
Uncertainty
Return Loss
35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
E6003A, E6003B,
E6004A, E6008B,
E6012A, E6013A _______ m 5 m _______ m