Test V. E6006A Power Meter Submodule D Single-Mode/Multimode Module Performance Tests
E6000C Mini-OTDR Users Guide, E0302
265
Performance Test for the Agilent E6000C with E6006A Power Meter submodule Page 2 of 2
Model Agt. E6006A submodule Report No. __________ Date ___________
Uncertainty/Accuracy test at Reference Conditions (Reference setting of power level 10.00
µW)
Wavelength Minimum Spec
(-3.6% of Ref.)
E6006A, DUT
Measurement
Results
Maximum Spec
(+3.6% of Ref)
Measurement
Uncertainty
1310.00 nm
1550.00 nm
9.640
µW
9.640
µW
_______
µW
_______
µW
10.360
µW
10.360
µW
_______
µW
_______
µW
Uncertainty/Accuracy Test: Wavelength 1310 nm
8156A
setting
81524A
Reference
Measurement
Minimum Spec
(-5% of Ref - 0.5
nW)
E6006A, DUT
Measurement
Results
Maximum Spec
(+5% of Ref +
0.5 nW
Measurement
Uncertainty
0 dB
5 dB
15 dB
25 dB
35 dB
45 dB
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________ W
________ W
________ W
________ W
________ W
________ W
Uncertainty/Accuracy Test: Wavelength 1550 nm
8156A
setting
81524A
Reference
Measurement
Minimum Spec
(-5% of Ref - 0.5
nW)
E6006A, DUT
Measurement
Results
Maximum Spec
(+5% of Ref +
0.5 nW
Measurement
Uncertainty
0 dB
5 dB
15 dB
25 dB
35 dB
45 dB
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________
µW
________
µW
________
µW
________
µW
_________nW
_________nW
________ W
________ W
________ W
________ W
________ W
________ W
Optional Test: Wavelength 850 nm
8156A
setting
81520A
Reference
Measurement
Minimum Spec
(-10% of Ref -
2.5 nW)
E6006A, DUT
Measurement
Results
Maximum Spec
(+10% of Ref +
2.5 nW
Measurement
Uncertainty
n/a ________
µW ________ µW ________ µW ________ µW ________ W
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6000C and is the answer not in the manual?

Agilent Technologies E6000C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6000C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals