Input the Injection Time in 24 hour format (example 1:00 PM as 1300) and press ENTER.
Input the Counting Period, in seconds, for measurements. Press the F4 [OK] key. You will be
prompted to count the following samples in the Well.
Reference Standard Standard supplied with the kit.
Room Background Remove all samples and sources from the Well.
Whole Blood Sample 4 ml of whole blood withdrawn from the patient five to ten
minutes post injection.
Plasma Sample 4 ml of plasma from the whole blood withdrawn from the
patient five to ten minutes post injection.
Perform each measurement according to the instructions in CHAPTER 14: LAB TESTS:
SETUP.
After the plasma sample counts have been accepted the RBC Volume (I125) Results Screen
(
Figure 16-8) will appear.
14:32
PLASMA VOLUME
DONE
PATIENT, ONE
PLASMA: 5237. ml 75. ml/kg
BLOOD: 9879. ml 141. ml/kg
RBC: 4642. ml 66. ml/kg
P
CALCULATED HEMATOCRIT: 47.0 %
Figure 16-8
Results are calculated as follows:
CountsPlasm
a
Net
4000 Counts Standard ReferenceNet
= (ml) Volume Plasma
×
CountsBloodNet Whole
4000 Counts Standard ReferenceNet
= (ml) Volume Blood Whole
×
Volume Plasma - Volume Blood = (ml) Volume RBC
November 07 LAB TESTS: BLOOD TESTS 16-7