The system also allows you to request the Auto Cal procedure, even when it is not indicated
by the Test function.
Press the HOME key, then the UTIL key. The main Utilities menu (
Figure 6-7) will appear.
UTILITIES
1. MULTICHANNEL ANALYZER
2. STAFF BIO ASSAY RESULTS
3. SETUP
4. DIAGNOSTICS
0. RETURN TO PROBE SCREEN
14:32
Figure 6-7
Note: Selection 0 will display “RETURN TO WELL SCREEN” if the Utilities menu is
accessed from the Well Main Screen.
Press 3 to select SETUP. The message “PASSWORD:” will appear. Type the last three digits
of your CAPTUS
®
600 serial number, and press the ENTER key. The Setup menu (Figure 6-
8
) will appear.
1. Ci/Bq
2. TEST DAT
A
3. THYROID UPTAKE
4. BIO ASSAY
6. ACTIVITY CONVERSION FACTORS
SETUP
5. AUTO CALIBRATION
7. WIPE AND MCA SETUP
8. OTHE
R
0. RETURN TO UTILITIES MENU
14:32
Figure 6-8
Press 5 to select AUTO CALIBRATION. The Detector Select Screen (
Figure 6-9) will appear.
6-6 ACCEPTANCE & QUALITY ASSURANCE TESTS November 07