CAPINTEC, INC. CAPTUS
®
600
1. WELL
2. PROBE
0. RETURN TO PREVIOUS SCREEN
14:32
MDA TEST
Figure 6-14
To test the Well, press
1. To test the Probe, press 2.
After Well or Probe is chosen, the Nuclide Entry Screen (
Figure 6-15) will appear.
13:06
MDA TEST
OK
NUCLIDE NAME: >TC99M
DONE
WELL
Figure 6-15
If another nuclide is desired, type in the new nuclide name (use the
* for the m in metastable
nuclides such as Tc99m).
Press
F4 [OK] to continue with the MDA Test. Figure 6-16 will appear.
Press
F1 [DONE] to abort the MDA Test.
6-10 ACCEPTANCE & QUALITY ASSURANCE TESTS November 07