CAPINTEC, INC. CAPTUS
®
600
Press F1 [NUM] to switch to numeric input and display “LET”.
Press F1 [LET] to switch back to letter input and display “NUM”.
When the protocol name is correct, press F4 [OK]. The Calculation Method Screen (
Figure
7-13
) will appear.
14:32
TU PROTOCOLS
METH
OK
NAME: PROT 1
CALCULATION: ONE CAPSULE
TU Calculation may be:
One Capsule
Standard
Two Capsules
Figure 7-13
Press F1 [METH] to change the Calculation method from “ONE CAPSULE” to “STANDARD
to “TWO CAPSULES”.
Refer to the CHAPTER 2: FUNCTIONAL & TECHICAL DESCRIPTION; SECTION:
THYROID UPTAKE PROTOCOLS for a description of the three methods.
When the desired calculation method is displayed, press F4 [OK].
Figure 7-14 will appear.
14:32
TU PROTOCOLS
NUCLD
OK
NAME: PROT 1
CALCULATION: ONE CAPSULE
NUCLIDE: I 131
NUMBER OF UPTAKES: 1
COUNTING REPETITIONS: 2
NUMBR
REPS
Figure 7-14
This screen is used to select the nuclide, number of uptakes and number of counting
repetitions: