CAPINTEC, INC. CAPTUS
®
600
Press F1 [NUCLD] to select the nuclide. The nuclide will cycle from “I131” to “Tc99m” to
I123” and back to “I131”.
Press F2 [NUMBR] to select the number of uptakes in the study. The choices are “1”, “2”, “3
or "4”.
Note: It's better to have too many measurements than too few measurements in a protocol.
If the physician decides, for example, that a patient does not need a third or fourth
measurement, the study can still be completed on the CAPTUS
®
600 without any
problem. But if the patient is assigned to a protocol that only has two measurements,
and then the physician decides that the patient needs a third measurement, there's
no way to add another measurement for this patient in the CAPTUS
®
600 system. If
you encounter that type of problem, the best solution is to make a printout of all
patient data, and then change the last measurement of the protocol to obtain results
of the additional measurement.
Press F3 [REPS] to select the number of repetitions for each measurement. The choices are
1” or “2”.
Note: If 2 repetitions are chosen, it is still possible to perform the measurement without
making the second repetition.
When all entries are correct, press F4 [OK].
Figure 7-15 will appear allowing the Pre-Dose
Measurement option and Counting Period to be set.
14:32
TU PROTOCOLS
OK
NAME: PROT 1
CALCULATION: ONE CAPSULE
COUNTING PERIOD :> 120 SEC
PRE-DOSE MEASUREMENT: NO
PRE
Figure 7-15
Input the desired Counting Period. The minimum is 10 seconds and the maximum is 9999
seconds.
Press F1 [PRE] to select the Pre-Dose Measurement option. This key toggles between “NO
and “YES”. The default is No.
When all entries are correct, press F4 [OK].
7-8 THYROID UPTAKE SETUP November 07