CAPINTEC, INC. CAPTUS
®
600
If the Calculation Method is One Capsule or Two Capsules, the Multi-Capsules option is now
chosen as shown in
Figure 7-16.
14:32
TU PROTOCOLS
OK
NAME: PROT 1
CALCULATION: ONE CAPSULE
MULTI-CAPSULES: NO
MULT
Figure 7-16
Press F1 [MULT] to switch from “NO” to “ONE” to “EACH” to “ALL”.
Refer to the CHAPTER 2: FUNCTIONAL & TECHICAL DESCRIPTION; SECTION THYROID
UPTAKE PROTOCOLS for a description of Multi-Capsules.
When the correct option is displayed, press F4 [OK].
The Measurement Times and Normal Value Range can now be input (
Figure 7-17) for each
uptake.
14:32
TU PROTOCOLS
OK
NAME: PROT 1
CALCULATION: ONE CAPSULE
INIMUM NORMAL VALUE : 0 %
AXIMUM NORMAL VALUE : 99 %
EASURE AT :>2 HOURS
UPTAKE NUMBER: 1 / 4
Figure 7-17
The Uptake Number indicates which uptake the data is for and the number of uptakes that
have been chosen for the study. In the example, this is the first of 4 uptakes.
At the “MEASURE AT” prompt, input the number of hours between dose administration and
this uptake measurement.
November 07 THYROID UPTAKE SETUP 7-9