Input the User ID number (reference the USER ID section on page 10-8) and press ENTER
to acknowledge the usage message.
If the calculated ratio is within limits and is useable for more than 12 hours, Figure 10-19
Mo99/Tc99m with Volume Entry screen will appear.
0.058 µCi/mCi
VOLUME? ml
_
Figure 10-19 Mo99/Tc99m with Volume Entry
Input the volume of the elution in ml and press ENTER. The concentration of the elution is
calculated and displayed as shown in Figure 10-20 Elution Concentration screen.
Note: The minimum value that can be input is 1.000. The maximum value that can be input
is 100.0.
1.232 Ci
123.2mCi/ml
Any Key to Continue
Figure 10-20 Elution Concentration
Press any key (except HOME) to print (if a printer is attached to the system) a ticket with the
details of the assay.
The prompt “PRINT AGAIN” will appear. If YES is pressed, another ticket is printed. If NO is
pressed, the Tests Menu will appear.
March 15 MEASUREMENT PROCEDURES 10 - 15