CAPINTEC, INC CRC
®
-25R
Select OTHER. Input the password (last 3-digits of the readout serial number) and press
. The Other Menu will appear.
ENTER
A4 - 4 APPENDIX IV April 12
Select SOURCES. The Test Source
s Menu will appear.
Select Cs137. The system will display
If the Test Source data has been input in the system, verify that these are the dates from
your Cs137 Test Source and press YES. Press NO if the dates are incorrect or if no source
data has been entered in the system.
Input the serial number of the Test Source and press ENTER. Input the Calibration Day and
press ENTER. Input the Calibration Activity and press ENTER. Press YES to answer yes to
“Use Daily?”.
The system will display the entered data. Verify the data and confirm with YES. To correct
any errors, press NO.
Press to return to the Measurement Screen.
HOME
Cs137
[displays Test
source data or
message “NO
SOURCE”]
OK? Y or N
1. Co57
2. Co60
3. Ba133
4. Cs137
5. Ra226
6. Constancy
1. User Keys
2. Sources
3. Moly Setup
4. Nuclides
5. Linearity
6. Remote / PC
For Technical Assistance, contact
Capintec’s only Authorized Service Center at 1-800-227-6832.