17.5.1 Frequency Analysis
To see help click on .
Frequency Analysis
Click on the Analysis type control to choose the analysis type. Each type
displays versus frequency. A connection diagram is given in the right hand
panel:
17.5.1.1 Measurements Group
Measurement Group
These types are used for making standard wideband, or swept frequency
narrow band measurements (if Use Swept Narrow Band is on).
RMS amplitude - RMS amplitude (in dBV or Volts)
Power - signal power (in dBm or Watts)
Power density- signal power density (in dBm/Hz or Watts/Hz)
Gain/Phase - Gain (in dB or G) on Channel A, and Phase (in degrees or
rads) on Chan B. Gain/Phase is best done Narrow Band.
To display in dB turn on the Amplitude/Power in dB Led in the Options
panel, for linear units turn the Led off.
17.5.1.2 Impedance, L, C Group
Impedance L/C Group
Uses the Circuit as selected (A, B or C) for the measurement setup. With this
group you can measure:
Impedance/Phase - |Z| in ohms or dBOhm on Chan A
Capacitance - C in F on Chan A
Inductance - L in H on Chan A
Chan B displays values based on the L-C Display drop down. These are:
Resr - |Effective Series resitsance| in ohms or dBOhm on Chan B
Q/D Factor - Quality or Dissipation Factor on Chan B
Phase - Phase in degreees or radians on Chan B