AS-410M Auto Slide Preparation System
1
AS-410M
Operation Manual
AUTO SLIDE PREPARATION SYSTEM
Chapter 1……….…………………….……Overview
Chapter 2…………………………… Installation
Chapter 3……………….……Preoperational Setup
Chapter 4……….….….. Operational Procedures
Chapter 5………..….. Errors and Troubleshooting
Chapter 6….….…Routine and Daily Maintenance
To use this product properly, please read this manual before use.
Please keep this manual close at hand for quick reference.
Ver. 1.4.24
IVD

Table of Contents

Questions and Answers

  • M
    Melanie GrayAug 14, 2025
    What to do if my DNS Industrial Equipment shows system check and initialization errors?
    • M
      Michael McleanAug 14, 2025
      If the DNS Industrial Equipment displays a 'System check and initialization errors', first, power the system OFF and then ON to retry the check. If the compressed air pressure is marked NG, ensure air is supplied at the specified pressure of 0.5 MPa (75psi). If 'System Doors CLOSED' is marked NG, check for and close any open doors.
  • C
    Colleen KellyAug 18, 2025
    What to do if I see a slide shortage caution on my DNS Industrial Equipment?
    • C
      Claire HaynesAug 18, 2025
      If the DNS Industrial Equipment displays a 'Slide shortage caution', it means that slides have run out during section slide preparation. To fix this, reload slides to resume and continue operation.
  • A
    averysheilaAug 21, 2025
    What does 'slide remaining caution' mean on my DNS Industrial Equipment and how do I fix it?
    • B
      Brenda Carpenter MDAug 21, 2025
      If the DNS Industrial Equipment displays a 'Slide remaining caution', it indicates that slides were left in the slide transfer area at the start of operation. Open the door and collect the slides from the slide transfer unit to resolve this.
  • K
    Kelsey OlsenAug 25, 2025
    Why does my DNS Industrial Equipment show a humidifying water low caution?
    • J
      Judy WardAug 26, 2025
      If the DNS Industrial Equipment gives a 'Humidifying water low caution', this means the water level in the humidifying water bottle is below the specified level. Check the humidifying water bottle and refill it if it's empty.
  • L
    Lisa GarrettAug 28, 2025
    What to do if I get a slide cassette error message on my DNS Industrial Equipment?
    • L
      Lisa StantonAug 28, 2025
      If the DNS Industrial Equipment displays a 'Slide Cassette error message', touch MAIN below the caution message to return to the 'MAIN' screen and select SLIDE PREPARATION. If the error is due to improper slide cassette loading, turn OFF the System power to correctly set slide cassettes on the racks.
  • M
    Michael JohnsonAug 31, 2025
    How to resolve an insufficient blade caution on DNS Industrial Equipment?
    • J
      jessica96Aug 31, 2025
      If the DNS Industrial Equipment shows an 'Insufficient blade caution', it means that the required number of replacement blades exceeds the number of the remaining blades. To resolve this, change the number of section slides to be prepared or the blade replacement conditions.
  • A
    awilliamsSep 4, 2025
    Why am I getting a carrier tape low caution on my DNS Industrial Equipment?
    • W
      wjonesSep 4, 2025
      If the DNS Industrial Equipment displays a 'Carrier tape low caution', it means the remaining length of carrier tape is too short for the specified number of section slides to be prepared. You can either change the number of section slides to be prepared or replace the carrier tape roll.
  • J
    Joseph OwensSep 7, 2025
    What to do if the slide collection box of my DNS Industrial Equipment is full?
    • D
      Dylan ChanSep 7, 2025
      If the DNS Industrial Equipment indicates that the slide collection box cannot accommodate the required number of slides, it is due to capacity limitations. Remove finished slides from the slide collection box and replace the slide collection box back in place.
  • R
    Richard Sharp Jr.Sep 12, 2025
    What to do if I get an input error on my DNS Industrial Equipment?
    • L
      Leslie ThomasSep 12, 2025
      If the DNS Industrial Equipment shows an 'Input error', make sure that you enter a valid value.

DNS AS-410M Specifications

General IconGeneral
BrandDNS
ModelAS-410M
CategoryIndustrial Equipment
LanguageEnglish