5-22 Infinity Vista VF6
5 ALARMS AND MESSAGES
Temperature (T)
Message Possible Cause Suggested Action
Message: T >UL
<value>
Message: T <LL
<value>
The patient's temperature falls
outside the current temperature
alarm limits.
• Observe the patient.
• Change the alarm limits for this
patient.
• If you suspect a temperature
probe failure, check and replace
it.
Message: MultiMed
Disconnected
Parameter value:
<blank>
The M
ULTIMED/NEOMED pod is
disconnected from the monitor.
• Verify the pod connections.
Message: T Fault
Parameter value: ***
The monitor has detected a
hardware failure.
• Turn the monitor off, then on.
• If the message does not clear,
take the unit out of operation and
call your Biomed.
Message: T
Unplugged
Parameter value: ***
The temperature probe is
disconnected from the M
ULTIMED/
NEOMED pod.
The M
ULTIMED/NEOMED cable is
defective.
The probe is defective.
The probe is not recommended for
use with your monitor.
• Verify the probe connection to the
pod.
• Verify the probe. If defective,
change the probe.
Message: T Test Fail The monitor has detected a
problem in the temperature sensor
during startup.
• If the test fails, turn the monitor
off, then on.
• If the message does not clear,
take the unit out of operation and
call your Biomed.
Message: T Too Low
Message: T Too High
Parameter value:
<blank>
The detected temperature is
outside the measurement range.
• Check your patient.
• Check the probe for defects.
• Check the patient cable
connections.
• Change the probe, if necessary.