5-22 Infinity Vista VF6
5 ALARMS AND MESSAGES
Temperature (T)
Message Possible Cause Suggested Action
Message: T >UL
<value>
Message: T <LL
<value>
The patient's temperature falls
outside the current temperature
alarm limits.
Observe the patient.
Change the alarm limits for this
patient.
If you suspect a temperature
probe failure, check and replace
it.
Message: MultiMed
Disconnected
Parameter value:
<blank>
The M
ULTIMED/NEOMED pod is
disconnected from the monitor.
Verify the pod connections.
Message: T Fault
Parameter value: ***
The monitor has detected a
hardware failure.
Turn the monitor off, then on.
If the message does not clear,
take the unit out of operation and
call your Biomed.
Message: T
Unplugged
Parameter value: ***
The temperature probe is
disconnected from the M
ULTIMED/
NEOMED pod.
The M
ULTIMED/NEOMED cable is
defective.
The probe is defective.
The probe is not recommended for
use with your monitor.
Verify the probe connection to the
pod.
Verify the probe. If defective,
change the probe.
Message: T Test Fail The monitor has detected a
problem in the temperature sensor
during startup.
If the test fails, turn the monitor
off, then on.
If the message does not clear,
take the unit out of operation and
call your Biomed.
Message: T Too Low
Message: T Too High
Parameter value:
<blank>
The detected temperature is
outside the measurement range.
Check your patient.
Check the probe for defects.
Check the patient cable
connections.
Change the probe, if necessary.