MESSAGES
VF6 Infinity Vista 5-21
Message: SpO
2
Searching
Parameter value:
<blank>
The monitor is analyzing the signal.
This occurs during the first 5 to 10 s
of operation or as a result of motion
artifact.
Wait until the message
disappears from the screen.
Message: SpO
2
Transparent
Message: PLS
Transparent
Parameter value: ***
The sensor may be disconnected
from the patient.
Sensor may be defective.
Check sensor and connections.
If problem persists, change
sensor.
Message: SpO
2
Unplugged
Message: PLS
Unplugged
Parameter value: ***
The sensors extension cable
disconnected from the pod.
The sensor is disconnected from
the extension cable.
The M
ULTIMED/NEOMED cable is
defective.
The sensor is defective.
The sensor is not recommended
for use with your monitor.
Verify sensor to extension cable
connection.
Verify extension cable to pod
connection.
Verify pod connection to monitor.
Verify correct sensor.
Message: Replace
SpO
2
Sensor
The SpO
2
sensor is not functioning
properly.
Replace the sensor.
Message: SpO
2
<LL
<value>
Message: SpO
2
>UL
<value>
Parameter value:
<value>
Message: PLS<LL
<value>
Message: PLS>UL
<value>
Parameter value:
<value>
The patient’s oximetry level or
pulse rate falls outside the current
alarm limits, or the alarm limits are
inappropriate for this patient.
The SpO
2
level has changed since
you last selected the alarm limits.
Elevated carboxyhemoglobin or
methemoglobin levels. Significant
levels of intravascular dyes.
Placement of sensor on an
extremity that has a blood pressure
cuff, arterial catheter or
intravascular line.
Faulty equipment.
Observe the patient.
Apply treatment if clinically
indicated.
Change the alarm limits.
If you suspect a sensor failure,
check and replace the sensor, if
necessary.
Message:
SpO
2
: non-Masimo
sensor
SpO
2
: non-Nellcor
sensor
Parameter Value: ***
Sensor not compatible with monitor
configuration.
Use compatible sensor (Masimo/
Nellcor).
Contact your hospital’s
Biomedical department to change
the monitor configuration.
Message:
SpO
2
: Duplicate
Device Connected
A MultiMed with an SpO
2
sensor
intermediate cable and an SpO
2
pod are connected to the monitor at
the same time.
Remove the sensor intermediate
cable from the MultiMed; or
Remove the SpO
2
pod.
Pulse Oximetry (SpO2) (continued)
Message Possible Cause Suggested Action