5-24 Infinity Vista VF6
5 ALARMS AND MESSAGES
Non-Invasive Blood Pressure (NBP)
Message Possible Cause Suggested Action
Message: NBP Cannot
Measure
Parameter value: ***
Faulty equipment.
Improper cuff placement.
Patient’s pulse too low.
Excessive patient movement.
Check the patient.
Check the hose and cuff.
Move the cuff to a limb with less
movement.
Check proper placement of the
cuff.
Restart measurement.
Message: NBP Artifact
Parameter value: ***
The patient is moving or shivering. Check the patient.
Move the cuff to a different limb
with less movement.
Message: NBP No
Pulsation
Parameter value: ***
Weak signal. Monitor is unable to
detect a sufficient number of
pulsations of adequate amplitude
within two minutes.
Check the patient and treat if
necessary.
Check the hose and cuff.
Check for proper size, placement
of cuff.
Message: NBP Cuff
Leak
Parameter value: ***
The monitor has detected a
significant drop in the cuff pressure
during inflation.
Check for leaks in the cuff or
hose.
Change the cuff and hose, if
necessary.
Message: NBP Fault -
Power Cycle Monitor
Parameter value: ***
Faulty equipment.
Improper cuff placement.
Transducer needs calibration.
NBP hardware needs repair or
calibration.
Hardware safety timer has expired.
Check the hose and cuff.
Check proper placement of cuff.
Turn the monitor off, then on if
you detect a hardware failure.
If the message does not clear,
take the unit out of operation and
call your Biomed or
DrägerService.
Message: NBP Mean
Only
Parameter Value:
NBP M <value>
NBP S <***>
NBP D <***>
The patient's pulse is too low for
the monitor to derive the systolic
and diastolic pressure values but
large enough to report the mean
pressure value.
Improper cuff placement
Check the patient.
Check proper placement of the
cuff.
Message: NBP Meas.
Time-out
Parameter value: ***
Measurement time has exceeded 2
minutes.
Repeat the measurement.
Message: NBP Open
Line
Parameter value: ***
There is no significant increase in
cuff pressure during the inflation
cycle.
Verify the cuff connection.