MESSAGES
VF6 Infinity Vista 5-25
Message: NBP
blocked Line
Parameter value: ***
There is a significant increase in
cuff pressure during the inflation
cycle.
Blocked output condition detected.
Cuff or tubing may be obstructed.
• Substitute cuff or tubing if
necessary.
Message: NBP
Overpressure
Parameter value: ***
The cuff pressure exceeds the
overpressure threshold.
• Call your Biomed.
Message: NBP Sys
<LL <value>
Message: NBP Sys
>UL <value>
Message: NBP Dia
<LL <value>
Message: NBP Dia
>UL <value>
Message: NBP Mean
>UL <value>
Message: NBP Mean
<LL <value>
Parameter value:
<value>
The patient’s pressure falls outside
the current alarm limits or these
limits are inappropriate for this
patient.
• Observe the patient.
• Change the current alarm limits
for this patient.
Message: NBP Out of
Range
Parameter value: ***
NBP values are above/below the
NBP measuring range.
• Check patient and treat, if
necessary.
• Verify that the inflation mode is
appropriate for the patient.
Change if necessary.
Message: NBP Low
Inflation Limit
Parameter value: ***
The patient’s systolic pressure is
above the current inflation limit.
• Change to a higher inflation limit.
Message: NBP Check
Cuff Size
Parameter value: ***
The wrong size NBP cuff is being
used on the patient.
• Verify the cuff size and placement
are proper for the patient.
• Verify all connections.
• Change the cuff if necessary.
Message: NBP Check
Hose Connection
Parameter value: ***
There is no significant increase in
cuff pressure during the inflation
cycle.
• Verify all connections.
• Change the hose if necessary.
Non-Invasive Blood Pressure (NBP) (continued)
Message Possible Cause Suggested Action