D-12 Infinity Vista VF6
D TECHNICAL DATA
Interfering Substances Carboxyhemoglobin may erroneously increase readings. The
level of increase is approximately equal to the amount of
carboxyhemoglobin present. Dyes, or any substance
containing dyes, that change usual arterial pigmentation may
cause erroneous readings.
Temperature
Input Connector 7-pin connector on MultiMed pod
Measurement Range 0 qC to +50 qC (32 qF to 122 qF)
Measurement Accuracy ±0.1 °C for a range of 0 °C to 50 °C
Probe Accuracy ±0.1 °C for a range of 0 °C to 50 °C
Total System Accuracy ±0.2 °C for a range of 0 °C to 50 °C
Average Response Time < 2.5 seconds
Linearity Linearity calculations to 0.02 qC by the software algorithm.
Excitation Signal DC source
Power to Probe < 50 PW (causing a potential self-heat error factor
of < 0.05 qC)
Test Signals – 5 qC and 50 qC ±0.1 qC with nominal thermistor
Filter Characteristic Low-pass filtering to minimize noise pickup.
Error Detection > 50 °C: high temperature, possibly caused by a shorted
thermistor.
–3 °C: low temperature, possibly caused by an open
thermistor.
Self-test failure
Degree of Protection Against
Electric Shock
CF
Defibrillation Protection In accordance with IEC 601-1A2
Non-Invasive Blood Pressure (NBP)
Parameter Display Systolic, Diastolic, Mean
Measuring Method Oscillometric technique
Modes of Operation Manual (single measurement), Continuous (5 minutes), or
Interval
Pulse Oximetry (SpO
2
) with Masimo SET Pod (continued)