OVERVIEW
VF6 Infinity Vista D-13
Interval Times 2, 2.5, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 30, 45, 60, 120,
180 and 240 min
Measuring Range
(Adult - 270mmHg)
Heart rate: 30 - 240 bpm
Systolic NBP: 30 - 250 mmHg
Mean NBP: 20 - 230 mmHg
Diastolic NBP: 10 - 210 mmHg
Measuring Range
(Pediatric - 180mmHg)
Heart rate: 30 - 240 bpm
Systolic NBP: 30 - 170 mmHg
Mean NBP: 20 - 150 mmHg
Diastolic NBP: 10 - 130 mmHg
Measuring Range
(Neonatal - 140mmHg)
Heart rate: 30 - 240 bpm
Systolic NBP: 30 - 130 mmHg
Mean NBP: 20 - 110 mmHg
Diastolic NBP: 10 - 100 mmHg
Default Inflation Pressure Adult (270): 160 mmHg ±10 mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Inflation Pressure After a Valid
Measurement
Adult (270): Previous SYS + 25 mmHg (±10)
Pediatric (180): Previous SYS + 25 mmHg (±10)
Neonatal (140): Previous SYS + 30 mmHg (±10)
Inflation Pressure
After a Technical Alarm
Adult (270): 160 mmHg ±10 mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Maximum Inflation
Pressure
Adult (270): 265 mmHg ±5 mmHg
Pediatric (180): 180 mmHg ±10 mmHg
Neonatal (140): 142 mmHg ±10 mmHg
Minimum Inflation
Pressure
Adult (270): 110 mmHg ±10 mmHg
Pediatric (180): 90 mmHg ±10 mmHg
Neonatal (140): 70 mmHg ±10 mmHg
Maximum Measurement Time Adult (270): 2 min
Pediatric (180): 2 min
Neonatal (140): 90 sec
(60s French homologation)
Maximum Measurement Time
Including a Retry
Adult (270): 3 min
Pediatric (180): --- (n/a)
Neonatal (140): --- (n/a)
Software Safety Cut-Off Adult (270): 273 ±3 mmHg
Pediatric (180): 215 ±3 mmHg (SWi)*
185-212 (SWh)*
Neonatal 140): 150 ±3 mmHg
Hardware Safety Cut-Off Adult (270): 300 ±30 mmHg
Pediatric (180): 300 ±30 mmHg
Neonatal (140): 158 ±2 mmHg
Static Cuff Accuracy ±3 mmHg
Calibration Range Adult and Pediatric: 10 - 260 mmHg ±3 mmHg
Neonatal: 10 - 150 mmHg ±3 mmHg
Non-Invasive Blood Pressure (NBP) (continued)