D-14 Infinity Vista VF6
D TECHNICAL DATA
Degree of Protection Against
Electric Shock
Type CF
Defibrillation Protection per EN 60601-2-30 (IEC 601-2-30)
*SWi refers to instant measurements; SWh refers to measurements lasting more than 5 seconds and less than
15 seconds.
Invasive Blood Pressure (IBP1, IBP2)
Accessories Transducers with a resistance of 300 to 2000: and an
equivalent pressure sensitivity of 5 PV/V/mmHg ±10%
Input Connector 7-pin connector on monitor
Number of Channels 2 max (requires IBP options)
Sweep Speed 25 mm/s
Measured Parameters Systolic, Mean and Diastolic blood pressure
Measurement Range -50 mmHg to 399 mmHg
Measurement Accuracy ±2 mmHg or ±3% (whichever is greater) after successful zero
and calibration (exclusive of transducer).
Resolution 1 mmHg or 0.1 kPa
Response Time d35 s (to reach 90% of the change of the lowest pulse rate of
25 1/min.
Zero Balance Range ±190 mmHg
Zero Resolution 0.24 mmHg
Zero Accuracy ±0.48 mmHg
Bandwidth DC to 16 Hz
Defibrillation Protection In accordance with IEC 601-2-34
Degree of Protection Against
Electric Shock
CF
Non-Invasive Blood Pressure (NBP) (continued)