GETTING S TARTED
VF6 Infinity Vista 2-7
STEPS: Charging Lead Acid Batteries with a Battery Charger
You can charge up to four additional lead acid batteries with the battery charger
available from Dräger. To do so:
1. Place the lead acid batteries on the battery charger.
2. Keep the lead acid batteries in the charger for at
least four hours to ensure a full charge.
STEPS: Removing a Lead Acid Battery from the Monitor
1. Connect the monitor to AC power or turn the monitor off.
2. Place the monitor on a flat surface.
3. Open the battery compartment door (back of the monitor).
4. Lift the battery from under the locking tab and pull the battery out.
CAUTION: Use the battery charger to charge lead acid batteries only. Lithium-ion
batteries are not compatible with the Dräger lead acid battery charger
NOTE: If the monitor is connected to AC power, monitoring can continue while you
replace the battery. If the monitor is not connected to AC power, turn the monitor off
before removing the battery in order to assure a proper monitor shut-down.
WARNING: To avoid explosion, do not disassemble, or
dispose of the battery in fire. Do not change the battery in the
presence of explosive hazards because of the possibility of
sparks. If the monitor is not to be used for a prolonged time,
remove the battery from the battery compartment.