2-6 Infinity Vista VF6
2 MONITOR SETUP
5. Insert the left side of the battery door into the hinges, and swivel the door
closed until the locking tab snaps into place.
6. Press the
Power fixed key to turn the monitor on. The green light indicator in
the key should light up.
7. Wait until the end of the self-tests and verify that the battery gauge appears at
the bottom of the display.
To prolong battery life, we recommend the following:
z Use the AC adapter whenever possible.
z Keep the battery charged.
STEPS: Charging the Battery with the Monitor
The battery that is delivered with a new monitor needs to be fully charged before
monitoring. When connected to line power via the AC Adapter, the monitor
automatically charges the battery and the green Battery Charger LED lights up (the
monitor can be turned on or off).
To charge the battery:
1. Connect the monitor to line power and let the lead acid battery charge for 5½
hours, the lithium ion battery for 8 hours. (The Battery Charging message
appears intermittently.)
2. Wait until the battery is fully charged before monitoring a patient. (The
Battery Charging message no longer appears.)
NOTE: The battery gauge appears only when you operate the monitor on battery
power.
NOTE: If the battery does not charge properly, the message Battery Charger Error
appears. Contact your Biomedical technician to replace the battery.