5-16 Infinity Vista VF6
5 ALARMS AND MESSAGES
Message: V Tach
Parameter value: VT
n or more consecutive PVCs have
been detected with a beat-to-beat
rate greater than or equal to the VT
rate set in the Alarm Tables. (n =
VT count parameter)
Observe the patient and treat if
clinically indicated.
Message: Bradycardia
Parameter value:
BRDY
8 or more consecutive QRS
complexes have been detected
with a beat-to-beat rate less than
80% of the patient’s normal rate.
Observe the patient and treat if
clinically indicated.
Message: RUN
Parameter value: RUN
3 to N-1* consecutive PVCs with a
beat-to-beat rate >= the VT rate
Observe the patient and treat if
clinically indicated.
Message: AIVR
Parameter value:
AIVR
3 or more PVCs with a rate < the
VT rate
Observe the patient and treat if
clinically indicated.
Message: SVT
Parameter value: SVT
N* or more consecutive normal
beats with a beat-to-beat rate >=
the SVT rate
Observe the patient and treat if
clinically indicated.
Message: CPT
Parameter value: CPT
A sequence of beats with the
pattern: normal, PVC, PVC, normal
Observe the patient and treat if
clinically indicated.
Message: BGM
Parameter value: BGM
A sequence of beats with the
pattern: normal, PVC, normal,
PVC, normal
Observe the patient and treat if
clinically indicated.
Message: TACH
Parameter value:
TACH
N* or more consecutive normal
beats with a beat-to-beat rate >=
the TACH rate
Observe the patient and treat if
clinically indicated.
Message: PAUS
Parameter value:
PAUS
A sequence of two normal beats
with an N-N* interval > the Pause
rate times the average N-N interval
(±100ms)
Observe the patient and treat if
clinically indicated.
Message: ST <lead>
Off
ST<lead1> and
ST<lead2> parameter
value: ***
The monitor has detected a lead-off
condition for the currently
processed lead. The cause could
be one of the following:
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on the electrode
Wrong cable type selection.
The M
ULTIMED/NEOMED pod may
be defective.
Check the patient cable and lead
wires carefully.
Check the MULTIMED pod.
Replace pod if defective.
Replace any cable or lead wire
that is suspect.
Reapply gel on the electrode or
change the electrode (if
disposable).
Select another ECG lead for
processing if the electrode or lead
cannot be replaced.
Heart Rate, Arrhythmia, ST Segment Analysis (continued)
Message Possible Cause Suggested Action