MESSAGES
VF6 Infinity Vista 5-17
Message: Cannot
Analyze
ST<lead1> and
ST<lead2> parameter
value: <blanK>
ST cannot analyze with a paced
beat
ST segment analysis cannot be
performed on pacemaker patients
Message: ST Too High
ST<lead1> and
ST<lead2> parameter
value: +++
Message: ST Too Low
ST<lead1> and
ST<lead2> parameter
value: ---
The detected ST value is outside
the measurement range.
Check the patient and apply
treatment if necessary.
Check the electrode placement
and change their position if
necessary.
Message: ST > UL
ST<lead1> and
ST<lead2> parameter
value: <value>
Message: ST < LL
ST<lead1> and
ST<lead2> parameter
value: <value>
The ST rate falls outside the
current upper or lower alarm limits.
The current alarm limits are
inappropriate for this patient.
Observe the patient carefully.
Apply treatment if clinically
indicated.
Change the alarm limits as
described in the “Cardiac
Monitoring” chapter.
Message: ST Lead
<1> and/or ST Lead
<2> Invalid
ST<lead1> and/or
ST<lead2> parameter
value: <blank>
Invalid ST<leadname>.
ST<lead1> or ST<lead2>
connection restored from a fault.
Excessive signal noise.
Bad contact of electrodes.
Interference from auxiliary
equipment.
Select correct ST lead(s).
Verify ST lead(s) connection.
Heart Rate, Arrhythmia, ST Segment Analysis (continued)
Message Possible Cause Suggested Action