5-18 Infinity Vista VF6
5 ALARMS AND MESSAGES
Respiration
Message Possible Cause Suggested Action
Message: Apnea
Parameter value: Apn
The monitor is detecting an apnea
event that exceeds the apnea time
set in the menu.
Incorrect placement of the
electrodes.
Improper threshold.
Missed shallow breaths.
Check the patient. Treat the
patient, if clinically indicated.
If the absence of impedance
changes is due to the placement
of the electrodes, reposition them
until the message clears.
If apnea is due to an improper
threshold, set the threshold until
the message clears.
If breaths are missing, use the
manual respiration mode to
ensure the detection of shallow
breaths.
Message: MultiMed
Disconnected
Parameter value: ***
The M
ULTIMED/NEOMED pod is
disconnected from the monitor.
Verify the pod connections.
Message: Coincidence
Parameter value:
<value>
(Neonatal mode only)
The impedance changes due to
cardiac activity are being counted
as breaths.
Check the patient.
Check the electrode placement.
Message: Rsp Fault
Parameter value: ***
Technical problem. Turn the monitor off, then on.
If the problem persists, take the
unit out of operation and call
DrägerService.
Message: Rsp High
Impedance
Parameter value: ***
Impedance out-of-range.
Broken cables.
Dried out gel on the electrodes.
Poor skin preparation.
Check the cables for damage.
Check the electrodes.
Move the electrodes, if
necessary.
Message: Rsp Testing
Parameter
value:<blank>
Self-tests in progress.
No displayed value.
Wait until the end of the self-tests.
Message: Rsp > <UL
value>
Message: Rsp < <LL
value>
Parameter value:
<value>
The respiration rate falls outside
the current upper or lower alarm
limits.
The current alarm limits are
inappropriate for this patient.
Observe the patient carefully.
Apply treatment if clinically
indicated.
Change the alarm limits as
described in the “Respiration
Alarms” section.
Message: Rsp Artifact
Parameter value: ***
High frequency artifact in the
signal.
Large baseline shifts.
Patient movement.
TENS interference.
IV infusion pump interference.
Check the electrodes and
reposition them.
Find the source of interference
and remove it.
Replace the electrodes if
necessary.
Change the cable or lead wires.