IRECTION 5344303-100, REVISION 3 VIVID P3 SERVICE MANUAL
8-88 Section 8-13 - Loading Base Image Software
Section 8-13 Loading Base Image Software (cont’d)
7.) Select one of the options for loading the system. Select choice [a] to load the complete disk.
Refer to Figure 8-74 on page 8-88 .
8.) Press “Yes” or “No” to continue.
9.) System DVD will be loaded as shown in Figure 8-76 on page 8-88 .
Figure 8-74 Selection for loading the system
WARNINGWARNING
While the software install procedure is designed to preserve data, you should select
choice [b] to format disk C only.
Figure 8-75 Confirmation on loading the system
Figure 8-76 System CD loading