Operation - Calibration
Part No.: 4418309_Rev09 Fusion4 MSC-L
Honeywell Installation & Operation Manual 5 - 269
9. When the preset volume is loaded completely the Calibration . Arm
n . Stream n . Wizard screen appears, on which the actual volume
must be entered.
10. Enter the actual volume and then select <Confirm>.
The Calibration. Arm n . Stream n . Wizard screen appears, on
which the new calculated meter factor appears.