6-9-4 TESTING A RECEIVER USING DIGITAL SIGNALING FORMATS
This procedure is an example for testing a Receiver using Digital Signaling Formats. In
this procedure, an RF signal is modulated with a POCSAG message.
o Press GEN Test Mode Key to access RF Generate Operation Screen.
o Enter Receiver frequency in RF Field.
o Select desired Output Connector for output of signal. If Receiver is tested with
direct connection, connect Receiver Input Connector to selected Output Connector.
Otherwise, connect antennas to selected Output Connector and Receiver Input
Connector.
o Set Output Level to desired level.
o Select DATA Generator for Modulation Source.
o Set Modulation Type to FM.
o Set Deviation Level for 4.0 kHz or as required.
o Set DATA Generator Format for POCSAG or POCSAG/.
o Press F4 Config.
o Set Data Rate to required setting for Receiver.
o Set Capcode Range to Receiver Capcode.
o Set POCSAG Function for desired test function.
o Press F6 RETURN.
o Press F5 BURST.
o Verify Receiver receives signal and decodes POCSAG Message correctly.