Series 2280 Precision DC Power Supply Reference Manual Section 3: Functions and features
077085503 / March 2019 3-39
When you run a list, you cannot change the following settings from front panel or using SCPI
commands. Sending the following commands causes an error 210, "Not permitted with list enabled."
Source settings that cause an error
MENU > Source
Settings > V-Set
HOME > V-Set
[:SOURce[n]]:<function>[:LEVel][:IMMediate][:AMPLitude]
(on page 7-77)
MENU > Source
Settings > I-Limit
HOME > I-Limit
[:SOURce[n]]:<function>[:LEVel][:IMMediate][:AMPLitude]
(on page 7-77)
MENU > Source
Settings > Output
Delay
MENU > Source
Settings > Delay
Rise
MENU > Source
Settings > Delay Fall
:OUTPut:DELay:STATe (on page 7-56)
:OUTPut:DELay:RISing (on page 7-55)
:OUTPut:DELay:FALLing (on page 7-54)
MENU > Source
Settings > Slew Rise
MENU > Source
Settings > Slew Fall
[:SOURce[n]]:VOLTage:SLEW:RISing (on page 7-79)
[:SOURce[n]]:VOLTage:SLEW:FALLing (on page 7-
80)
MENU > Source
Protect > Over
Voltage Protection
MENU > Source
Protect > Over
Current Protection
MENU > Source
Protect > Maximum
Voltage
HOME > Protect >
OVP
HOME > Protect >
OCP
HOME > Protect >
VMax
[:SOURce[n]]:<function>:PROTection[:LEVel] (on
page 7-76)
[:SOURce[n]]:VOLTage:LIMit[:AMPLitude] (on page
7-78)
MENU > Source
Delay > Source
Delay
MENU > Source
Delay > State
[:SOURce[n]]:DELay (on page 7-74)
[:SOURce[n]]:DELay:STATe (on page 7-75)