Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-43
:CALibration:PROTected:SAVE
This command saves the calibration to nonvolatile memory after the calibration procedure has been completed.
Usage
:CALibration:PROTected:SAVE
Details
If the instrument exits the calibration mode by the following command without first saving the new
constants, the previous constants are restored.
:CALibration:STATe OFF
Example
Save calibration constants to memory.
Also see