Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-125
Example
:SYST:PRES
Set the instrument to restore the factory default settings
and erase all user data stored in nonvolatile memory,
such as list parameter and setup information.
Also see
*RST (on page 7-7)
Reset default values (on page 2-88)
:SYSTem:RWLock
This command locks out front-panel controls.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:RWLock
Details
When this command is sent, all of the front-panel keys are disabled. The instrument can only be
controlled using the remote interface.
Example
:SYST:RWLock
Disables all of the front-panel keys.
Also see
:SYSTem:LOCal (on page 7-120)