Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-140 077085503 / March 2019
:TRACe[n]:FEED
This command selects the source of readings.
Recall settings
Instrument reset
Power cycle
Usage
:TRACe[n]:FEED <name>
:TRACe[n]:FEED?
Channel number: For single-channel instruments, the value is 1
Select the source of readings:
ï‚§ Store the readings before the math calculation: SENSe
ï‚§ Store the readings after the math calculation: CALCulate
Details
When you select SENSe for the source, the readings stored in the trace buffer do not have a math
calculation even if the math state is ON. For details about data flow, refer to Displayed measurements
(on page 4-15).
Example
:TRACe:FEED SENS
:TRACe:FEED?
Set the source of readings to SENSe and query the
source.
Output is:
SENS
Also see
None