Series 2280 Precision DC Power Supply Reference Manual Appendix C: Calibration and adjustment
077085503 / March 2019 C-25
Calibration constants
You can use the following command to read all the calibration constants in order of their index
number:
:CALibration:PROTected:DATA?
For Model 2280S-60-3, the default calibration constants are listed in the following table.
Default
constant
tolerance
123, Step5 10A shunt value out of range;
128, Step7 10A shunt value out of range
116, Step3 100mA shunt value out of range;
120, Step4 100mA shunt value out of range
111, Step1 V-DAC zero DMM reading out of
range
112, Step1 V-ADC zero collected data error
113, Step2 V-DAC full scale DMM reading out
of range
114, Step2 V-ADC full scale collected data
error
122, Step5 I-ADC 10A zero DMM reading out of
range
124, Step5 I-ADC 10A zero collected data
error
125, Step5 I-ADC 1A zero collected data
error
127, Step7 I-ADC 10A full scale DMM reading
out of range
129, Step7 I-ADC 10A full scale collected
data error
126, Step6 I-DAC zero collected data error
130, Step8 I-DAC full scale collected data
error
115, Step3 I-ADC 100mA zero DMM reading out
of range
117, Step3 I-ADC 100mA zero collected data
error
118, Step3 I-ADC 10mA zero collected data
error
119, Step4 I-ADC 100mA full scale DMM
reading out of range
121, Step4 I-ADC 100mA full scale collected
data error