Series 2280 Precision DC Power Supply Reference Manual Section 2: General operation
077085503 / March 2019 2-43
5. Power on the instrument and repeat steps 1 through 3.
The following figure shows the error that is displayed on the front panel.
Figure 28: Open lead error
You can retrieve the error number and message through a remote interface by sending
the :SYSTem:ERRor? command.
Reversed sense leads
When the following connections occur, the instrument generates a reversed sense leads error.
Sense HI is connected to Output LO
Sense HI is connected to Sense LO
Sense LO is connected to Output HI
When a reverse sense leads error occurs, the instrument:
Generates an overprotection error message.
Displays an error dialog box on the screen.
Turns off the output.
Resets the voltage setting value and current limit to defaults.
Sets the digital I/O line 3 to logic high.