Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-79
[:SOURce[n]]:VOLTage:SLEW:RISing
This command specifies and queries the rising rate for the voltage source.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
100 V/s
Usage
[:SOURce[n]]:VOLTage:SLEW:RISing <value>
[:SOURce[n]]:VOLTage:SLEW:RISing?
[:SOURce[n]]:VOLTage:SLEW:RISing? DEFault
[:SOURce[n]]:VOLTage:SLEW:RISing? MINimum
[:SOURce[n]]:VOLTage:SLEW:RISing? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
The voltage rising rate: 10 V per second to 1000 V per second.
Details
Slew rate is defined as the maximum rate of change of output voltage per unit of time. The slew rising
rate is applied to any voltage increase.
When you turn on the output, the voltage rises at the specified rate.
Example
:VOLT:SLEW:RIS 20
Set the voltage rising rate to 20 V per second.
Also see
[:SOURce[n]]:VOLTage:SLEW:FALLing (on page 7-80)