Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-120 077085503 / March 2019
:SYSTem:LOCal
This command takes the instrument out of remote operation and restores the operation of front-panel controls.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:LOCal
Details
When the instrument is in remote operation, all the front-panel keys except the LOCAL and OUTPUT
keys are locked out. When this command is sent, the instrument restores operation of the front-panel
controls. However, you can send SYSTem:LOCal to take the instrument out of remote and restore
operation of front panel controls.
If the trigger source or arm source is set to manual triggering, the TRIGGER key is enabled when the
instrument is in the remote mode.
Example
:SYST:LOC
Returns to local mode and you can control the
instrument from the front panel.
Also see
:SYSTem:RWLock (on page 7-125)