EasyManuals Logo

Keithley 2280 Series Reference Manual

Keithley 2280 Series
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #259 background imageLoading...
Page #259 background image
Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-58 077085503 / March 2019
:OUTPut:PROTection:TRIPped?[<channel>]
This command queries whether overvoltage, overcurrent, overtemperature, or reverse fault protection has been
tripped.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:OUTPut:PROTection:TRIPped?
<channel>
The channel number (default value is CH1):
ï‚§ Channel 1: CH1
ï‚§ Channel 2: CH2
ï‚§ Channel 2: CH2
The
Details
There are six types of responses, including:
OCP indicates over-current protection occurred.
OVP indicates over-voltage protection occurred.
OTP indicates over-temperature protection occurred.
RVS indicates reverse sense fault detected.
OPS indicates open sense error protection occurred.
NONE indicates nothing occurred.
Example
:OUTP:PROT:TRIP? CH1
If the overvoltage protection circuit in channel 1 is
tripped, the query returns OVP.
If the protection circuit is not tripped, the query returns
NONE.
Also see
None

Table of Contents

Other manuals for Keithley 2280 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280 Series and is the answer not in the manual?

Keithley 2280 Series Specifications

General IconGeneral
BrandKeithley
Model2280 Series
CategoryPower Supply
LanguageEnglish

Related product manuals