Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-11
The Series 2280 has the following overlapped commands:
ï‚§ *TRG
ï‚§ :INITiate
ï‚§ :CALibration:PROTection:STEP1:DATA
ï‚§ :CALibration:PROTection:STEP2:DATA
ï‚§ :CALibration:PROTection:STEP3:DATA
ï‚§ :CALibration:PROTection:STEP4:DATA
ï‚§ :CALibration:PROTection:STEP5:DATA
ï‚§ :CALibration:PROTection:STEP6:DATA
ï‚§ :CALibration:PROTection:STEP7:DATA
ï‚§ :CALibration:PROTection:STEP8:DATA
Also see
CALibration subsystem (on page 7-34)
*OPC (on page 7-5)
*TRG (on page 7-9)
:ABORt[n]
This command stops all trigger model commands on the instrument.
Usage
:ABORt[n]
Channel number: For single-channel instruments, the value is 1
Details
When this command is received, the instrument stops the trigger model.
With continuous initiation disabled, the trigger model of the specified channel goes into the idle state.
With continuous initiation enabled, operation continues at the top of the trigger model. For details,
refer to :INITiate[n]:CONTinuous (on page 7-51).
Also see
Trigger model (on page 3-25)