Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-16 077085503 / March 2019
:MEASure[n]:<function>?
This command makes a measurement and returns the reading.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:MEASure[n]:<function>?
:MEASure[n]:<function>? <range>
:MEASure[n]:<function>? <range>,<resolution>
[n]
Channel number: For single-channel instruments, the value is 1
<function>
The measurement function to which this setting applies:
Concurrent: CONCurrent[:DC]
Current: CURRent[:DC]
Voltage: VOLTage[:DC]
<range>
The range value.
Current range and concurrent range:
10 mA range: 0.01
100 mA range: 0.1
1 A range: 1
10 A range: 10
Voltage range:
Model 2280S-32-6: 32 V
Model 2280S-60-3: 60 V
<resolution>
4.5 digit resolution: 4
5.5 digit resolution: 5
6.5 digit resolution: 6
Details
This command makes a measurement using the specified function and stores the reading in a
reading buffer.
If the buffer is set to NEVer mode, this command only makes a measurement and returns the
reading without saving in the buffer.