Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-16 077085503 / March 2019
:MEASure[n]:<function>?
This command makes a measurement and returns the reading.
Usage
:MEASure[n]:<function>?
:MEASure[n]:<function>? <range>
:MEASure[n]:<function>? <range>,<resolution>
Channel number: For single-channel instruments, the value is 1
The measurement function to which this setting applies:
ï‚§ Concurrent: CONCurrent[:DC]
ï‚§ Current: CURRent[:DC]
ï‚§ Voltage: VOLTage[:DC]
The range value.
Current range and concurrent range:
ï‚§ 10 mA range: 0.01
ï‚§ 100 mA range: 0.1
ï‚§ 1 A range: 1
ï‚§ 10 A range: 10
Voltage range:
ï‚§ Model 2280S-32-6: 32 V
ï‚§ Model 2280S-60-3: 60 V
ï‚§ 4.5 digit resolution: 4
ï‚§ 5.5 digit resolution: 5
ï‚§ 6.5 digit resolution: 6
Details
This command makes a measurement using the specified function and stores the reading in a
reading buffer.
If the buffer is set to NEVer mode, this command only makes a measurement and returns the
reading without saving in the buffer.