Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-148 077085503 / March 2019
Example
TRAC:TRIG:VOLT 20
TRAC:TRIG:VOLT?
The value of the voltage threshold is set to 20.
Also see
:TRACe[n]:TRIGger:VOLTage:STATe (on page 7-148)
:TRACe[n]:TRIGger:VOLTage:DIRection (on page 7-147)
:TRACe[n]:TRIGger:VOLTage:STATe
This command determines whether the voltage level trigger function is enabled.
Usage
:TRACe[n]:TRIGger:VOLTage:STATe <state>
:TRACe[n]:TRIGger:VOLTage:STATe?
ON or 1: Enable the voltage level trigger function
OFF or 0: Disable the voltage level trigger function
Example
TRAC:TRIG:VOLT:STAT 1
TRAC:TRIG:VOLT:STAT?
The voltage level trigger function is enabled.
Also see
:TRACe[n]:TRIGger:VOLTage[:LEVel]
:TRACe[n]:TRIGger:VOLTage:DIRection (on page 7-147)