Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-146 077085503 / March 2019
:TRACe[n]:TRIGger:OCCur?
This command queries whether a level trigger has occurred.
Usage
:TRACe[n]:TRIGger:OCCur?
Example
Queries whether a level trigger has occurred.
If a level trigger has occurred, the value of 1 will be
returned.
Also see
None
:TRACe[n]:TRIGger:OFFSet
This command defines the number of points at which the instrument begins to log data in the buffer.
Usage
:TRACe[n]:TRIGger:OFFSet <value>
:TRACe[n]:TRIGger:OFFSet?
Example
TRAC:TRIG:OFFS 10
TRAC:TRIG:OFFS?
10 points before the level trigger activation are not cleared.
Also see
None