Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-30 077085503 / March 2019
:CALCulate2:FORMat:ELEMents
This command specifies or queries data elements to be included in the data string. It only influences minimum
and maximum query results.
Recall settings
Instrument reset
Power cycle
Usage
:CALCulate2:FORMat:ELEMents <itemList>
:CALCulate2:FORMat:ELEMents <itemList>,<channel>
:CALCulate2:FORMat:ELEMents?
:CALCulate2:FORMat:ELEMents? <channel>
The available data elements:
ï‚§ Voltage and current reading: READing
ï‚§ Reading units: UNIT
ï‚§ Reading number: RNUMber
ï‚§ Timestamp: TSTamp. For example, output might be: 07/01/2013
19:21:36.2553
The channel number (default value is CH1):
ï‚§ Channel 1: CH1
Channel 2: CH2
Details
This command affects the output of the following commands:
ï‚§ :CALCulate2:IMMediate?
ï‚§ :CALCulate2:DATA?
You can specify one or more elements. <itemList> is a string that must be enclosed in single or
double quotes. Each element in the list must be separated by a comma (,).