Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-20 077085503 / March 2019
:ARM[:SEQuence[n]]:SOURce
This command selects the control source for the arm layer.
Recall settings
Instrument reset
Power cycle
Usage
:ARM[:SEQuence[n]]:SOURce <source>
:ARM[:SEQuence[n]]:SOURce?
Channel number: For single-channel instruments, the value is 1
The trigger source for the arm layer:
ï‚§ Immediate triggering; no delay or gating in arm detection layer:
IMMediate.
ï‚§ External triggering: EXTernal
ï‚§ Manual triggering; the trigger signal is generated by pressing the
TRIGGER key: MANual
ï‚§ Bus triggering by the command *TRG: BUS
Example
Select bus triggering as the trigger source
for the arm layer.
Also see
:TRIGger[:SEQuence[n]]:SOURce (on page 7-151)
Trigger model (on page 3-25)