EasyManuals Logo

Keithley 2280 Series Reference Manual

Keithley 2280 Series
457 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #352 background imageLoading...
Page #352 background image
Series 2280 Precision DC Power Supply Reference Manual Section 7: SCPI command reference
077085503 / March 2019 7-151
:TRIGger[:SEQuence[n]]:SOURce
This command selects the control source for trigger layer.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
IMM
Usage
:TRIGger[:SEQuence[n]]:SOURce <source>
:TRIGger[:SEQuence[n]]:SOURce?
[n]
Channel number: For single-channel instruments, the value is 1
<source>
The trigger source for trigger layer:
ï‚§ Immediate triggering: IMMediate; no delay or gating in the
detection layer
ï‚§ External triggering: EXTernal
ï‚§ Manual triggering: MANual; the trigger signal is generated by
pressing the TRIGGER key
Example
:TRIG:SOUR EXT
Select external triggering as the trigger source.
Also see
:ARM[:SEQuence[n]]:SOURce (on page 7-20)
Trigger model (on page 3-25)